Xray Photoelectron Spectroscopy (XPS) Omicron
Xray Photoelectron Spectroscopy  (XPS) Omicron
Make: Omicron Nano Technology GmbH
Model:
Year of purchase: 2010
Funded by: DST, Govt of India
Location: ICMS
Slot timings:
Contact:
Book New Request

  • Binding energy resolution down to approximately 0.4 eV may be obtained by using the monochromatized X-ray source.
  • Operating pressure ~ 2 x10-11 mbar.
  • Sample types: Thin films, polished solids, dispersed powders on substrates like silicon, copper film, and silver film. Powder samples duly adhesive to copper tape.
  • All samples should be dry.
  • Sample dimensions: X,Y (max) 10 mm, 10 mm max
  • Source: MgKa (1253.6 eV) and AlKa X-rays (1486.6 eV) anodes
  • Routine XPS analysis with wide scan of samples.
  • Auger spectroscopy, valance band.
  • Typical XPS probing depth of 2-10 nm.
  • Minimum probing area: 60 μm x 60 μm
  • Maximum probing area: 1.75 mm x 2.75 mm
Modes Duration (hrs)
M1 Max 5 elements
Sl. No. - Category Fees ( in Rs/- ) + additional 18% GST
01. JNCASR for M1 Free
04. Industry for M1 Rs 3000 (max 5 elements) 500 for every additional element
02. Other Academia for M1 Rs 1500 (max 5 elements) 300 for every additional elemen
03. Industry Partner for M1 Rs 3000 (max 5 elements) 500 for every additional element